Omron
HEM-7080IC *
Equivalent/Identical devices rules
Equivalent devices have not been validated but evidence has been provided and approved by the STRIDE BP Scientific Advisory Board showing that they are equivalent to a device that has fulfilled the criteria for validated devices. These devices have minor model differences which do not affect the blood pressure measurement accuracy, such as memory improvement, provision of bluetooth, PC connection, printer facilities, or additional functions. Identical devices have no difference from a STRIDE BP approved device, apart from model name change.
Market Availability
Not Available
Population
Adults
Children
Use
Home
Measurement Site
Upper arm cuff
Measurement Method
Automated oscillometric
Additional Features
Automated storage of multiple readings, Body movement detection, PC link
Validation studies of Omron 705IT (HEM-759-E), to which this device is equivalent
Population
Adults
Protocol and version (year)
ESH-IP 2002
Year of validation study publication
2003
Link
Population
Children
Protocol and version (year)
AAMI 1992
Year of validation study publication
2006
Link
Population
Children
Protocol and version (year)
ESH-IP 2002
Year of validation study publication
2006
Link
Population
Adults
Protocol and version (year)
BHS 1993
Year of validation study publication
2006
Link